Датчик процесса Keyence FD-XS1

COM_VIRTUEMART_PRODUCT_NOSTOCK
ModelFD-XS1
Supported pipe materialsMetal pipes, Plastic pipes (soft/hard) *1
Supported fluidsLiquids (water, oil, adhesive, grease, chemical solution, etc.) *1
Supported fluid temperature0 °C (no freezing on the pipe surface) to 100 °C (Pipe surface temperature)
Clamp set modelPlastic pipe/tube attachmentFD-XC1R1, FD-XC1R2
Metal pipe attachmentFD-XC1M
Maximum rated flow rate0 to 1000 mL/min
Zero cut flow rate20 mL/min*2 (variable, default)
Display resolutionInstantaneous flow rate0.1/1/10 mL/min (Displayed on controller)
Shot amount0.001/0.01/0.1/1 mL (Displayed on controller)
RepeatabilityPlastic pipe /tube attachmentResponse time: 50 msF.S.±0.6 %*3*4
Instantaneous flow rate±6 mL/min*3*4
Response time: 500 ms±1.9 mL/min*3
Metal pipe attachmentResponse time: 50 msF.S.±1 %*3*4
Instantaneous flow rate±10 mL/min*3*4
Response time: 500 ms±3.2 mL/min*3
HysteresisVariable
Integrated unit display0.1/1/10/100/1000/10000 mL (displayed on controller)
Display methodStatus indicator
Environmental resistanceEnclosure ratingIP65/IP67 (IEC60529) , IP68G (JIS C0920) *5
Operating ambient temperature0 to 60 °C (No freezing)
Operating ambient humidity35 % to 85 %RH (No condensation)
Vibration resistance10 to 55 Hz, double amplitude 1.5 mm, 2 hours each for X,Y,Z direction
Shock resistance50 G 11 ms, 3 times each for X,Y,Z direction
MaterialSensor headHead body: PPS/PPSU, in-cable amplifier: PPS, cable: PVC, controller connector: PPS/PBT/POM
Clamp setFor plastic pipeBody, fixing screw: PPS, detection surface: special rubber, pipe support rubber: FKM, sensor head fixing screw: SUSXM7
For metal pipeMetal: SUS304/SUSXM7, detection surface: special rubber, clamp support rubber: FKM, sensor head fixing screw: SUSXM7
WeightApprox. 230 g
*1 Liquid must allow for the passage of an ultrasonic pulse, as well as not contain large air pockets or excessive bubbles. Readings may become unstable depending on the type of pipe.
*2 The zero cut flow rate can be changed in the settings. When using the unit with a low flow rate range, perform an origin adjustment when the fluid is not moving if you change the zero cut flow rate.
*3 This specification is valid when the flow velocity distribution is stable. This value does not take into account the effects of pulsation or fluctuations in flow velocity distribution due to facility factors. Convert the F.S. (full scale value) listed in the table according to the rated flow range.
*4 The longer the response time is set, the more repeatability is improved. As a guideline, use v (50 ms/response time) times.
*5 The connector part of the sensor head cable is IP65/IP67.